Philips - Respironics Trilogy 202 Portable design for intra-hospital transport, The Trilogy 202 is both a volume-control and pressure-control ventilato. A year ago Upload service manual How to upload a Trilogy 202 service manual for the benefit MedWrench community? Reply: 4 Replies. MedWrench User Guide. Trilogy 202 offers all the conventional volume-control modes: AC, SIMV (with or without PS), and CV. The Trilogy 202 is designed to compensate for leaks in a volume control mode. This means consistent tidal volume delivery even during noninvasive or when cuff leaks are present.
|
|
Philips Respironics Trilogy 202 Manual
How to upload a Trilogy 202 service manual for the benefit MedWrench community?
4 RepliesTrilogy Respironics Manual
We are trying to complete a PM on a 202. Every single ventilator we have tried passes every step up to step 55, which every vent fails at. 55 gives an error 8017 to 0030.0550, Negative flow verify.. dP2 at -0.0 (setpoint 140) in S LPM. We also fail several other pressure test verify levels after that, about 13 afterwards. WHAT ARE WE DOING WRONG?? Could somebody please tell us what we're doing wrong? The Respironics guy that is supposed to be guiding us thru this mess is UNRESPONSIVE and we can't get thru to anyone else at Respironics. HELP!!
1 ReplyI need to perform a PM on this unit. It is a new piece of equipment to me.